Input type: PT20/PT100/PT200/PT500/user-defined thermal resistor B/R/S/N/K/J/T/E thermocouple/thermopile
Number of input channels: 2
Main functions: sensor matching to achieve ultra-high precision measurement.
Maximum and minimum values, average value, maximum-minimum value display.
Over-limit alarm.
℃/F/K/Q/mv measurement unit switching.
Data storage up to 390,000 records, including time information.
USB data directly connected to computer.
Power supply: 6~15VDC external power supply or 4 AAA batteries.
Communication rate: 9600bps (JT-11)
1200/2400/4800/9600/19200/38400bps can be set (JT-11A)
Purpose:
①Precision resistance detection, platinum resistance temperature corresponding resistance detection saves the consumption of second-class standard thermometers.
②Constant temperature field acceptance.
Resistance measurement accuracy (AT25 ℃):
Range
|
Sampling mode
|
Resolution (typical) Ω
|
Error (Max) Ω
|
300Ω
|
Ω
|
0.0001
|
±(0.0005+0.0006%*READ)
|
TRUEQ
|
3kQ
|
Ω
|
0.006
|
±(0.01+0.005%*READ)
|
TRUEQ
|
DC voltage measurement accuracy (AT25 ℃)
Range mV
|
Resolution (typical) μV
|
Error (Max) μV
|
70
|
0.1
|
±(1+0.004%*READ)
|
Resistance measurement accuracy (AT25 ℃): Temperature Drift
Project
|
Impact per 1 ℃ change of environment (maximum)
|
Resistance
|
2ppm*READ
|
Voltage
|
7ppm*READ
|
Resistance measurement accuracy (AT25 ℃): electrical performance
The parameter conditions given in this table are: ambient temperature 25 ℃, power supply voltage 6V, dual-channel resistor sampling, and measuring range 300 Ω.
Project
|
Condition
|
Typical value
|
Working current
|
Turn on backlight
|
13.5mA
|
Turn off backlight
|
11mA
|
Resistance measurement sampling time
|
Step change> 3mΩ, two channels simultaneously measure resistance
|
High resolution
|
Low resolution
|
6s
|
2s
|
Step change <3mΩ, two-channel simultaneous measurement of resistance
|
High resolution
|
Low resolution
|
18s
|
6s
|
Voltage measurement sampling time
|
Two-channel simultaneous measurement of mv
|
0.5s
|
Thermocouple measurement sampling time
|
Simultaneous measurement of thermocouple signals with two channels
|
High resolution
|
Low resolution
|
6.5s
|
2.5s
|
Resistance measurement accuracy (AT25 ℃): the overall accuracy after matching with the sensor
Enter model number
|
Range (℃)
|
Accuracy (℃)
|
PT100
|
-50 ~ 200
|
± 0.01
|
PT50
|
-60 ~ 400
|
± 0.02
|
T-Type thermocouple
|
-200 ~ -50
|
± 0.3
|
-50 ~ -300
|
± 0.15
|
N-type thermocouple
|
0 ~ 400
|
± 0.3
|
300 ~ 600
|
± 0.5
|
S-type thermocouple
|
0 ~ 600
|
± 0.5
|
600 ~ 1100
|
± 1.5
|