Input type: 2/3/4-wire PT100 and tungsten-rhenium/B/R/S/N/K/J/T/E thermocouples
Output mode: 0~10V (adjustable). RS485 (optional)
Output resolution: 1mv
Data update cycle: 0.5 seconds
Thermocouple cold end compensation accuracy (AT25℃): (maximum) ±0.1℃
Operating environment: -30~70℃ 0~95% relative humidity
Special function: Can be matched with the sensor to achieve ultra-high precision measurement. When thermocouple input, an external temperature compensation element can be connected to achieve instant compensation. Sensor open circuit indicator alarm.
Electrical isolation: power-input-output/RS485
Isolated from each other, isolation voltage 600V
Power supply: 18~36VDC (other ranges can be customized)
Accuracy (AT25 ℃ & AT0.5 maximum sampling rate):
Enter model number
|
Range (℃)
|
A/D conversion accuracy
|
D/A conversion accuracy
|
Impact of environment change of 1 ℃ (temperature drift)
|
PT100
|
-200~500
|
± 0.05 ℃
|
±0.05% of range
|
0.003 ℃
|
B-Type thermocouple
|
300~1820
|
± 1.40 ℃
|
±0.05% of range
|
0.030 ℃
|
S-type thermocouple
|
0 ~ 1768
|
± 0.65 ℃
|
±0.05% of range
|
0.025 ℃
|
N-type thermocouple
|
-200~1300
|
± 0.35 ℃
|
±0.05% of range
|
0.019 ℃
|
K-type thermocouple
|
-180~1372
|
± 0.25 ℃
|
±0.05% of range
|
0.018 ℃
|
J-type thermocouple
|
-180~760
|
± 0.20 ℃
|
±0.05% of range
|
0.006 ℃
|
T-Type thermocouple
|
-200~400
|
± 0.20 ℃
|
±0.05% of range
|
0.009 ℃
|
E-type thermocouple
|
-50 ~ 1000
|
± 0.17 ℃
|
±0.05% of range
|
0.006 ℃
|
Digital/analog conversion
|
|
|
|
0.002% Span
|
Note 1: The digital/analog conversion accuracy and its temperature drift given in the table are for the output range of 0~10v.
Other range calculation methods: digital/analog conversion accuracy = 0.05*10/range; Digital/analog conversion temperature drift = 0.002*10/range.
For example, 1~5V output: digital/analog conversion accuracy = 0.05*10/(5-1)= 0.125%
Note 2: Total digital accuracy of thermocouple measurement: analog/digital conversion accuracy +0.1 ℃ (cold end compensation accuracy)
Overall accuracy after using sensor matching function (transmitter + sensor)
Enter model number
|
Range (℃)
|
A/D conversion accuracy (℃)
|
PT100
|
-50 ~ 200
|
± 0.05
|
S-type thermocouple
|
0 ~ 650
|
± 0.70
|
650~1100
|
± 1.50
|
N-type thermocouple
|
0 ~ 650
|
± 0.40
|
650~1100
|
± 1.90
|
T-Type thermocouple
|
-200 ~ -70
|
± 0.40
|
-70 ~ 400
|
± 0.20
|
Note 1: The sensors listed in the table are the most commonly used, stable and representative sensors in each temperature range. For details of accuracy in a wider temperature range, please consult the manufacturer.
Note 2: The thermocouple matching function is a patented technology and must be completed before leaving the factory.
Note 3: The accuracy of thermocouples in the table includes the accuracy of cold junction compensation and does not need to be superimposed separately.